Polyethersulphone (PES) membrane Filter Capsules For fine, clarifying, biological filtration of liquids

Short Description:

Puno Pes capsule fifilters is family of full size capsule fifilters
with Staubli connection at the vent, which enables in-line
integrity test.
The PES membrane bio-burden reduction capsule utilizes
single layer hydrophilic polyethersulfone membrane. It offers
broad chemical compatibility, high flflow rate and low protein absorption rate

Polyethersulfone is particularly suited for the filtration of products that
contain substances that adsorb to the media The lower binding characteristics
of polvethersulfone make it a good choice for filtration of valuableprotein solutions


Product Detail

Video

Product Tags

Polyethersulphone (PES) membrane Filter Capsules For fine, clarifying, biological filtration of liquids,
5441307H4, 5445307H7, 5447307H1, 5447307H2, Gamma Capsules, Maxicaps, Midicaps Capsules,

Capsule Filter




Key Features
◇ Pleated structure improve filters dirt holding capacity and high throughput, and prolongs the filters service life;

◇ The disposable structure does not require a stainless steel housing, making the filters more economical, low cost and convenient than the conventional filter methods;

Capsule filters are using pleated processing, with compact structure and large filter area, applicable to small flow rate and large volume solutions filtration. Filter are sealed by means of melting, no glues and adhesives so don’t cause any pollution for the filter products. They will experience 100% integrity test, purified water washing, and pressure test before delivery. And there are various materials for selecting and using.

◇ Minimize product and filtrate residues, it’s a best choice for laboratories and small-sized mechanical terminal filter;

◇ Screw-based connection, making the filter easy replacement and convenient, multiple types of connectors are available for selection;

Ordering Information

 

 

 

No.

Filter medium

No.

Removal rating (μm)

No.

Filter area

No.

Packing quantity

P

PP

001

0.01

10

1000 cm2

1

1pcs

S

PES

002

0.02

20

2000 cm2

3

3pcs

F

PTFE

010

0.1

30

Others

6

6pcs

N

N66

020

0.2

 

 

 

 

S

Stainless steel

045

0.45

 

 

 

 

 

 

100

1.0

 

 

 

 

 

 

150

1.5

 

 

 

 

 

 

 

300

3.0

 

 

 

 

 

 

 

 

 

500

5.0

 

 

 

 

 

 

 

 

 

10H

10

 

 

 

 

 

 

High capacity platform for conventional sterile filtration
Application Industry:Pharma / Biotech
Biological Information Bacterial Retention
Quantitative Retention of 1 × E7 CFU/cm² Brevundimonas Diminuta ATCC® 19146 per ASTM methodology
Compliance Information
Bacterial Endotoxins
< 0.25 EU/mL as determined by the LAL test
Biological Reactivity
All materials of this filter element meet the requirements of the current USP Biological Reactivity tests <88> for plastics Class VI (Systemic Injection, Intracutaneous and Implantation tests).
Extractables
The total amount of extractables is well below the limits established by the current USP unter “Sterile Water for Injection”.
Connections (Physical)
Connector Inlet
Bayonet Adapter with 222 Double O-ring (27)
Connector Outlet
Bayonet Adapter with 222 Double O-ring (27)
Dimensions
Filter Size
20″(1.2 m² / 12.9 ft²)
Filtration Area
1.2 m²
Outer Diameter
70 mm
General Specifications
Capsule/Cartridge Format
Cartridges
Maximum Allowable Back Pressure
2 bar (29 psi) at 20°C; 1 bar (14.5 psi) at 80°C
Maximum Diffusion
18 mL/min at 2.5 bar | 36 psi at 20 °C per 10 inch element
Minimum Required Bubble Point
3.2 bar | 46 psi
Vent Valve
N/A
Materials of Construction
Housing Material
Polypropylene (PP)
Material o-ring | seal
Silicone (SI)
Support Fleece
Polypropylene (PP)
Operating Conditions
Maximum Differential Pressure
20 °C: 5 bar; 80 °C: 2 bar
Physicochemical Information
Non-Fiber Releasing
This filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
Pore Size Final-filter
0.2 µm
Pore Size Pre-filter
0.45 µm
Product Information
Brand
Sartopore® 2
Sterility
Sterilization Method
Autoclavable & In-Line Steam Sterilizable
Usage Protocols
Sterilization Procedures
In-Line Steam Sterilization: at 134 °C, 20 min., 0.5 bar, min. 25 Cycles
Autoclaving: at 134°C, 30 min, 2 bar, min. 25 Cycles
Related Products


  • Previous:
  • Next:

  • Related Products